SB
 
Filter
Filtering allows the instrument to ignore unwanted mechanical vibrations or sudden impacts to the scale. Correctly set filtering shortens the time required to output a stable weight value and can increase the accuracy of the reading.
It is suggested to run your system while making adjustments to WAVERSAVER® to create the actual working conditions for the scale system (ie: turn on motors, mixers, augurs, etc. that may affect the scale).
WAVERSAVER® and Num Averages should be set as low as practical for your application to optimize response time.
Gross
Live Weight Reading

The filters can be set with PRESET Combinations of WAVERSAVER® and Num Averages, or CUSTOM with any valid selection of WAVERSAVER® and Num Averages. Go to Step 1 with PRESETs or directly go to Step 2 with CUSTOM values.

PRESET

Check ONE preset Checkbox. After making your selection, click Submit to see how your selection performs.
OFF
FAST
BALANCED
STABLE
CUSTOM
[NOTE: Only one Checkbox should be Checked]

CUSTOM
Step 1 -
Select a WAVERSAVER® setting that stabilizes the Live Weight Reading shown. After making your selection, click Save Parameters to see how your selection performs.
WAVERSAVER®

Step 2 -
Set a Num Averages value to minimize the effects of impact to the scale. Usually a setting of 10 is enough to clear small bumps to the scale and product impact events.
Num Averages
WAVERSAVER+ Settings
Variation Threshold
Motion Threshold

Step 3 -
Save your parameters.
Once you have saved your parameters, you are done with this section. Return to the Main Wizard Page
Previous Page
Back

Home - Calibration - Diagnostics - Monitor - Setup - Wizard - HI 6800 Series Online Support Site

Hardy Process Solutions
San Diego, California, USA
Tel. +1 (858) 278-2900 ▪ (800) 821-5831
Fax +1 (858) 278-6700
www.hardysolutions.comhardyinfo@hardysolutions.com
hardysupport@hardysolutions.com

homecalibrationdiagnosticsMonitorindexWizardhardy